WEBVTT 00:00:00.400 --> 00:00:01.920 Welcome to this guide 00:00:02.040 --> 00:00:05.120 which sets out the correct use of the Movymia Pen. 00:00:05.400 --> 00:00:08.480 The Movymia Pen is made up of the following parts: 00:00:08.880 --> 00:00:11.920 pen cap, cartridge holder, housing, 00:00:11.920 --> 00:00:16.040 flange, threaded rod, display window, 00:00:16.040 --> 00:00:19.120 push botton and dosage knob. 00:00:19.120 --> 00:00:21.720 The needle consists of the outer needle cap, 00:00:21.720 --> 00:00:24.480 inner needle cap, the pen needle itself 00:00:24.480 --> 00:00:26.440 and the peel foil. 00:00:26.440 --> 00:00:30.440 The Movymia cartridge completes the unit. 00:00:33.280 --> 00:00:39.200 We will start by showing you how to use your Movymia Pen for the very first time. 00:00:44.480 --> 00:00:47.120 Start by removing the pen cap. 00:00:49.400 --> 00:00:53.600 If you’re changing a cartridge, remove the empty cartridge first. 00:00:55.240 --> 00:00:58.480 Now insert the new Movymia cartridge into the cartridge holder, 00:00:58.760 --> 00:01:00.600 metal cap first. 00:01:07.480 --> 00:01:11.800 Using your finger, carefully push back the threaded rod in a straight line 00:01:11.800 --> 00:01:13.560 as far as it will go. 00:01:14.600 --> 00:01:15.320 Next, 00:01:15.400 --> 00:01:17.920 attach the cartridge holder to the pen housing 00:01:18.040 --> 00:01:22.080 by rotating it 90° clockwise until it stops. 00:01:24.720 --> 00:01:27.800 Now attach the needle to the pen. 00:01:28.560 --> 00:01:30.960 Start by pulling off the peel foil. 00:01:34.720 --> 00:01:38.120 Now screw the needle clockwise onto the cartridge holder 00:01:38.480 --> 00:01:43.760 making sure that the needle is attached correctly and firmly to the cartridge holder. 00:01:44.960 --> 00:01:47.520 Remove and keep the outer needle cap 00:01:48.080 --> 00:01:51.160 then remove and discard the inner needle cap. 00:01:56.720 --> 00:02:01.400 After insertion of a new cartridge, we do what is called “priming” of the cartridge. 00:02:01.720 --> 00:02:04.080 This is also a functionality test of the pen 00:02:04.160 --> 00:02:07.680 that is carried out before the first injection from a new cartridge. 00:02:08.280 --> 00:02:12.600 Turn the dose knob clockwise until you see a droplet sign on the dose display. 00:02:13.240 --> 00:02:16.320 Make sure that the two indicator stripes are aligned. 00:02:18.480 --> 00:02:22.160 During dose dialling, the pen will make an audible clicking sound 00:02:22.360 --> 00:02:24.600 and you will feel some resistance. 00:02:25.400 --> 00:02:27.320 Press the button down fully 00:02:27.440 --> 00:02:29.840 and keep it pressed until the dose indication 00:02:29.840 --> 00:02:32.040 has returned to the start position. 00:02:33.640 --> 00:02:36.720 A few drops of liquid should emerge from the needle tip. 00:02:37.120 --> 00:02:42.120 aIf no drops emerge, repeat this last step until the drops emerge 00:02:42.120 --> 00:02:46.440 (please note that this step should not be repeated more than 4 times). 00:02:48.800 --> 00:02:53.040 The pen is now primed and ready for use as part of your daily routine.